General requirements for CiGas-NFs
NFs (i.e. observational platforms) have to comply to the following general requirements:
- Collocated measurements of at least 6 different VOCs and NOx
- Accompanied measurements of meteorological parameters in high time resolution near the place of the air intake port for the analysers. (wind speed, wind direction, air temperature, and air pressure)
- Use of a standard traceable to the CiGas Measurements
- The performance of the measurement instrument has to be checked against a target gas in regular intervals (e.g., monthly).
- Reported data include the measured amount fractions, precisions and uncertainties as specified in SOP-VOC (2014) and SOP-NOx (2014), as well as flags
- Data have to be regularly (e.g., monthly) reviewed for consistency with existing data from the same measurement site and against similar European measurement sites à @VOC@-Tool (https://voc-qc.nilu.no/QATool)
- Data are submitted with metadata and data flagging at least yearly (by end of March).
- Data is reviewed by CiGas and will be discussed at an annual data quality meeting
- Participation in round-robin exercises
- Accept and organize performance audits by CiGas.
General requirements for exploratory platforms are under development.
The ACTRIS Documentation on technical concepts and requirements for ACTRIS Observational Platforms can you find here.
Measurement Guidelines
For NOx and VOC can be found here SOP for NOx and VOC
Labelling step 1a
The workflow to proceed in labelling step 1a for reactive trace gas insitu measurements can be found here CiGas labeling step 1a Workflow
The link to the labelling interface can be found here https://actris-nf-labelling.out.ocp.fmi.fi/
Trainings/tutorials related to CiGas activities
EBAS tutorials for data submission:
CiGas-Empa @VOC@ tutorials:
- https://polybox.ethz.ch/index.php/s/eO2ifEby7an5sLo
- https://polybox.ethz.ch/index.php/s/uI9WdN81AluL7gO
Services provided to NFs
CiGas is providing services to users on demand. Requests should be addressed through SAMU. Services available:
Service |
Instrument |
Modality of access |
Provision | |
---|---|---|---|---|
Access to measurement guidelines (MG) and standard operation procedures (SOP) |
NOx by Chemilumenescence NMHCs by Thermodesorption-Gaschromatography |
Virtual access | ||
Data quality assurance workshop |
NOx by Chemilumenescence NMHCs by Thermodesorption-Gaschromatography |
Virtual/physical access | ||
Training | ||||
Hands-on training | ||||
Round robin | ||||
On-site audit | ||||
Virtual audit | ||||
Intercomparison | Condensables - CI-APi-TOF | physical access | ||
Instrument calibration | Condensables - CI-APi-TOF | physical access |